How to calibrate a TNM 40AX metallographic mounting machine if the heating temperature is inaccurate?

If the heating temperature of the TNM-40AX metallographic mounting machine is inaccurate, professional temperature calibration equipment can be used for calibration.
First, accurately connect the sensor of the calibration equipment to the temperature measurement point of the mounting machine, then set different temperature values, let the mounting machine heat to the set temperatures, and record the actual temperatures displayed by the calibration equipment.
Compare the difference between the set temperature and the actual temperature, and adjust the temperature control system of the mounting machine according to the difference, such as modifying the temperature control parameters. Repeat the above steps multiple times until the temperature error is within the allowable range.